gonyx:registration_reliability_is_red_or_yellow
Differences
This shows you the differences between two versions of the page.
Next revision | Previous revision | ||
gonyx:registration_reliability_is_red_or_yellow [2012/07/12 08:24] – created mkroeber | gonyx:registration_reliability_is_red_or_yellow [2013/02/28 12:48] (current) – external edit 127.0.0.1 | ||
---|---|---|---|
Line 1: | Line 1: | ||
- | ====== Registration Reliability Display is red or yellow when pins are set correctly ====== | + | ====== Registration Reliability Display is red or yellow when pins seem to be set correctly ====== |
The registration reliability display is a security feature. It's mandatory to move the blue boxes exactly | The registration reliability display is a security feature. It's mandatory to move the blue boxes exactly | ||
Line 17: | Line 17: | ||
- Now you can measure the pin distance in the axial view.\\ {{: | - Now you can measure the pin distance in the axial view.\\ {{: | ||
- | The horizontal length from Pin 2 to Pin 3 must be a multiple of 5mm. The length of the perpendicular line segment | + | **The horizontal length from Pin 2 to Pin 3 must be a multiple of 5mm. The length of the perpendicular line segment |
- | from the horizontal line to the center of Pin 1 must be 25mm. There might always be smaller deviations (~0.1 up to 0.5mm). | + | from the horizontal line to the center of Pin 1 must be 25mm.** There might always be smaller deviations (~0.1 up to 0.5mm). |
This is normal and based on several interference factors and the Doctors usually know this. But if there is a bigger deviation | This is normal and based on several interference factors and the Doctors usually know this. But if there is a bigger deviation | ||
(above 1mm), the customer should start investigating, | (above 1mm), the customer should start investigating, |
gonyx/registration_reliability_is_red_or_yellow.1342081480.txt.gz · Last modified: 2013/02/28 12:44 (external edit)